Introduction

The AZ100 Zirconia Oxygen Analyzer is a versatile system designed primarily for the OEM boiler and burner controls market.

The system is a low-temperature type www.abb-drive.com designed to work in process temperatures up to 800 °C (1472 °F) and with a maximum sensor mounting flange temperature of 400 °C(752 °F).

The analyzer provides oxygen computation, with readout and retransmission, based on the probe mV output signal. The output signal (E mV) is Nernstian in form and follows the equation:

E (mv) = 0.0496T(log10 P0 / P1) ± CmV

Where:  T = Absolute temperature

P0 = Reference O2 partial pressure

P1 = Sample O2 partial pressure

C = Cell constant (mV zero offset)

0.0496 = Faraday’s gas constant

Probe design

The probe uses the proven and innovative ABB electrode and cell design technology which has been so reliable in other ABB zirconia probe designs.

The flexible probe design gives a range of intake tube lengths to suit all applications and an optional filter/flame arrester making it safe for use where groups IIB and IIC gases may occur in the process being measured. The probe has options of male thread NPT or BSPT mountings. As with all previous ABB designs of low temperature probes, the AZ100 probe is site-serviceable.

As the sensor housing is located on the outside of the duct wall, diffusion of reference air into the sensor housing is sufficient; thus eliminating the need for an air pump or instrument air supply.

The reference air diffuses into the housing through a porous membrane which restricts entry to air only and maintains the IP66 (NEMA 4X) protection.

Transmitter design

The transmitter has, as standard, high/low alarm relays and a single linear or logarithmic isolated retransmission. Display features include %O2. cell temperature, heater control output, cell mV, alarm set points, calibration sequence diagnostics and output settings.

At system startup the transmitter controls the level of power to the mains-powered heater within the probe to eliminate the risk of thermal shock to the sensor.

Based on the proven 4600 Series of transmitters, the AZ100 transmitters are environmentally protected to NEMA 4X (IP65)*, and meet the requirements EN61326 for industrial locations.

The AZ100 transmitters have a green, backlit LCD display and four tactile membrane switches for operation and programming. The measured value display is a 5-digit, 7-segment LCD, while the information display is a 16-character, single line, dot-matrix.

The information display can be user-programmed for display in English, French, German or Spanish.